Photometer Testing Guidlines
Photometer Testing Guidlines
|
|
•
|
Only use tablets designed for use in Lovibond® photometers. |
•
|
Vials, caps and stirring rods should be cleaned thoroughly after each analysis to prevent interferences. Even minor reagent residues can cause errors in the test result. |
•
|
Do not use fast dissolving/rapid tablets (green print on the foil) as you will get inaccurate results. |
•
|
The outside of the vial must be clean and dry before starting the analysis. Clean the outside of the vials with a towel. Fingerprints or other marks will be removed. |
•
|
Zero calibration and test must be carried out with the same vial as there may be slight differences in optical performance between vials. |
•
|
Store the unit in a cool dry place away from moisture. |
•
|
Do not store it in a plant room, service van or any place where chemicals are stored. |
•
|
Rinse test cells, test cell lid and crushing rod between each test to avoid cross contamination of samples. |
•
|
Do not leave wet sample cells upside down in the case. |
•
|
Do not shake the sample to dissolve tablet reagent, always use the crushing rod and place the test cell on a hard flat surface to avoid breaking the bottom of the cell. |
•
|
The vials must be positioned in the sample chamber for zeroing and test with the mark on the vial aligned with the mark on the instrument. |
•
|
Contamination of the lens in the sample chamber can result in errors. Check at regular intervals and – if necessary – clean the light entry surfaces of the sample chamber using a moist cloth or cotton buds. |
•
|
Always perform zeroing and test with the vial cap tightly closed. Only use the cap with a sealing ring. |
•
|
Bubbles on the inside wall of the vial lead to incorrect measurements. To prevent this, remove the bubbles by swirling the vial before performing the test. |
•
|
Avoid spillage of water into the sample chamber because this can lead to incorrect test results. |
•
|
Large temperature differences between the instrument and the environment can lead to errors – e.g. due to the formation of condensation in the area of the lens or on the vial. |
•
|
To avoid errors caused by stray light do not use the instrument in bright sunlight. |
•
|
Always add the reagent tablets to the water sample straight from the foil without touching them with the fingers. |
•
|
The reagents must be added in the correct sequence. |
•
|
We offer a photometer service that includes cleaning and calibration. |
•
|
Avoid using the user calibration feature. If calibration is in question phone for details. |